3-2 Startup and Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 NDIR
Controller Board and the network "self-install"
(bind together) during initial startup. If the
system contains more than one Analyzer
Module, the startup procedure will interrogate
the network to locate and identify all
components on the network. The user will
have to bind appropriate combinations of
components after the startup sequence.
After the warm-up period (about one hour for
the NDIR Analyzer Module), all modules are
completely functional.
Check the tune and detector signal values
against the factory settings listed in the
Diagnostic Service menus. If both settings
are within ±5% tolerance of factory setting, go
to Section 3-4 below for binding and Section
3-5 on page 3-4 for calibration. If not, refer
first to Section 4-11 on page 4-6 for
instructions about oscillator tune/source
balance shutter adjustments.
3-4 BINDING
To achieve full coordination between Analyzer
Modules and associated I/O Modules, the
user must bind those components together in
the System Set Up portion of the Technical
Configuration Menu in software.
Figure 3-1. Run Mode Display
Figure 3-2. Main Menu Display
Analyzer PQ 322-14
23.2 % CO
0
pp
m
50
Secondar
y
Variable: XXXX
Secondar
y
Variable: XXX
Secondar
y
Variable: XXXX
Secondar
y
Variable: XXXX
Display Parms. Menu Dual Info
F1
F2 F3 F4 F5
23.2 % CO Analyzer XXXXXXXX
Main Menu
Expert controls and setup ...
(Operational configuration)
Technical level configuration ...
(Diagnostic and manufacturing/service)
Display Parms. Info
F1
F2 F3 F4 F5
Basic Controls
Delete alarm message!