3
Section 1 Specifications
Specifications are subject to change without notice.
Component Description
Microprocessor-controlled measuring unit with measured value display,
temperature display, and menu-driven system
Controller Operating Temperature
–20 to 60 °C (–4 to 140 °F); 95% relative humidity, non-condensing with
sensor
load <7 W; 20 to 40 °C (–4 to 104 °F) with sensor load <25 W
Controller Storage Temperature –20 to 70 °C (–4 to 158 °F); 95% relative humidity, non-condensing
Enclosure NEMA 4X/IP66 metal enclosure with a corrosion-resistant finish
Power Requirements
100–230 V ac ±10%, 50/60 Hz;
Power: 11 W with 7 W sensor load, 35 W with 25 W sensor load
Pollution Degree/Installation Category II; II
Outputs
Two (Analog (4–20 mA)) outputs, maximum impedance 500 ohm. Optional digital
network connection. IrDA digital connection.
Relays
Three SPDT, user-configurable contacts rated 100–230 V ac, 5 Amp
resistive maximum
Controller Dimensions ½ DIN—144 x 144 x 150 mm (5.7 x 5.7 x 5.9 inches)
Controller Weight 1.6 kg (3.5 lb)