Series 4060 Installation
Teledyne Analytical Instruments 35
2. When the CRC closes, remove the span command.
When CRC opens again, the span calibration is done, and the sample
is being analyzed.
3.3.2.7
RANGE ID RELAYS
There are three dedicated Range ID relay contacts. They are
assigned to relays in ascending order—Low range is assigned to Range
1 ID, Medium range is assigned to Range 2 ID, and High range is
assigned to Range 3 ID. Table 3-6 lists the pin connections. Contacts
are normally open, and they close when 4060 switches to that particular
range.
Table 3-6: Range ID Relay Connections
Pin Function
21 Range 1 ID Contact
38 Range 1 ID Contact
22 Range 2 ID Contact
39 Range 2 ID Contact
19 Range 3 ID Contact
18 Range 3 ID Contact
34 Not Used
35 Not Used
3.3.2.8 NETWORK I/O
A serial digital input/output for local network protocol. At this
printing, this port is not yet functional. It is to be used in future options
to the instrument. Pins 13 (+) and 29 (–).
3.3.2.9 PIN OUT TABLE
The following table summarizes all the outputs/inputs available in
the 50 pin D-Sub connector on the back panel of the analyzer.