Teledyne 4060 TV Converter Box User Manual


 
Installation Series 4060
38 Teledyne Analytical Instruments
Figure 3-3: Gas Connections
Gas connections to the instrument are made at the 1/8”or 1/4”
stainless steel tube fittings provided on the rear panel. Note that the
Purge and Sensor Vent fittings are 1/4” while all other gas connections
are 1/8”.
It is recommended that all gas tubing leading to the connections on
the back of the analyzer be of the coiled type. This will facilitate sliding
the unit out of the case without disconnecting the gas supply to the
analyzer.
Before tubing is connected to the system, it must be decontaminated
to eliminate any hydrocarbon deposits. Using a small torch, heat each
length of tubing while passing nitrogen through it until it glows red.
Begin at the nitrogen source end and proceed down the length of the
tube, “chasing” the red glow (and hydrocarbon deposits) down to the
open end of the tube. Cap the tubing while not in use with suitable non-
contaminating caps.
All sample, calibration, and supporting gas lines, which deliver gas
to the analyzer, must be decontaminated before connection; vent lines do
not.
When connecting the various gas lines to the system, be absolutely
certain that no “dead ends” are left; that is, no unused branch lines
should be left capped off, where pockets might form of material that is