Series 4060 Operation
Teledyne Analytical Instruments 49
2. The MAIN MENU and SUB MENU screens have a flashing
arrow on the left side indicating which item is being pointed
to.
3. MODIFIABLE ITEM screens have a flashing value
somewhere in the line.
4.6.4.1 MAIN MENUS
The MAIN MENU consists of 15 functions you can use to customize
and check the operation of the analyzer. They are listed here with brief
descriptions:
SPAN: Set up and/or start a span calibration
ALT-SPAN: Set up and/or start an alternate span calibration.
MODEL: Displays Manufacturer, Model, and Software
version of the instrument.
SELF-TEST: The instrument performs a self-diagnostic
routine to check the integrity of the power supply, output
boards, cell and amplifiers.
ALARMS: Used to set the alarm setpoints and determine
whether each alarm will be active or defeated, HI or LO
acting, and failsafe or not.
RANGE: Used to set up three analysis ranges that can be
switched manually.
ANALOG-OUT ADJUST: Adjust the analog output with
offset and scaling.
LINEARIZATION: Linearization option.
CHANGE STREAM: (optional feature) Controls a relay
switch for using alternate gases. Has no effect during span
mode (span gas relay setting used at that time).
AUTO-CAL: Used to define and/or start an automatic
calibration sequence. AUTOCAL is an optional feature not
included in the standard configurations of the 4060.
TIMING: Low level application specific timing parameters.
Factory set. Do not change these.