3654 Portable H2/N2 Analyzer - Installation 9 of 66
Operator Manual
ORBISPHERE
1 Installation
This section provides necessary information to set up the instrument. If you have any
questions or experience any difficulties, do not hesitate to contact your Hach Ultra
representative regarding this procedure.
1.1 What you have Received
The ORBISPHERE 3654 portable analyzer includes the following major components:
• An instrument, model 3654/xxx, with keypad, display, and handle.
• A membrane-covered thermal conductivity (TC) sensor, either model 3127x for H
2
measurement or model 31570 for N
2
measurement.
• A flow chamber, either model 32013A with flow adjustment valve for N
2
analyzers,
or model 32015 without flow adjustment valve for H
2
analyzers.
• A WIN3654 program on a CD with a PC communications cable.
Also included with your shipment are two C-type batteries, a recharge kit in a plastic case
containing materials to maintain your sensor and a model 32824A refilling hose to refill
the purge gas cylinder
Locate the instrument close to the sample being analyzed, and to the Personal Computer
(PC) if one is being used. A LEMO-6 connector for RS-232 serial output to a PC is on the
right side of the instrument (see Fig 1-1 above).
Make sure you install a fully charged set of batteries, or connect the instrument to an
external power source (see “External Power Supply (optional)” on page 11), before
switching the instrument on and exposing the sensor’s membrane to any liquid.
Fig 1-1:
ORBISPHERE 3654 Portable Analyzer