22 of 66 Operating Instructions - 3654 Portable H2/N2 Analyzer
Operator Manual
ORBISPHERE
2.3.3 Viewing Stored Measurements
To return to the measurement mode, put the instrument into standby and then turn back
ON again without holding down any additional buttons.
2.4 Storing and Accessing Measurements From a Computer
If you have made measurements and stored them in the analyzer, you should be ready
to bring them into the WIN3654 program for viewing, copying, saving and printing. See
also “Options Setup” on page 27 for additional information on the WIN3654 program.
2.4.1 Downloading Stored Values
To download the stored results from the analyzer to the PC, choose the DownLoad data
command from the Logger menu.
The DownLoad window presents a display of the stored measurements from the
instrument. The window displays five columns of data:
Sample (sequence number of the sample)
Conc (concentration of the measured gas)
Date (date of the measurement)
Time (time of the measurement)
Sample Description
The descriptions can be modified for your applications using the procedures described
below.
1) Put the instrument into standby (by
pressing the POWER key)
2) Hold down the Up Arrow button
while switching the instrument back
ON. The LCD displays a sample
location number.
3) Scroll through the numbered
sample locations of all the stored
values using the Up Arrow and
Down Arrow buttons.
4) To view the actual gas
concentration measurement value
at a particular sample number,
press the STO button. The LCD
now displays the stored value for
that sample number.
5) Press STO a second time to return
to the next numbered location
display, to continue scrolling or
view another stored value.