1.8.6 Flow Chamber
The flow chamber inlet and outlet should be free of any obstructions. The model 32013A
flow chamber, for dissolved N
2
measurements, has a flow adjustment valve on the outlet.
The model 32015 flow chamber for H
2
measurements has no valve.
Dissolved Gas Measurements
Adjust the sample flow rate using the needle valve on the exit. When analyzing
carbonated beverages, sufficient pressure must be provided to avoid degassing of the
beverage during its passage through the flow chamber. Check for the absence of
bubbles in the inlet tubing.
Gaseous Measurements
Place the outlet tubing in a container of water, ensuring that the flow chamber outlet is
unobstructed (on the 32013A flow chamber, open the outlet valve completely). Then,
adjust the inlet sample flow valve (main supply) to obtain one bubble per second at the
outlet.