Emerson TCL TV Converter Box User Manual


 
16
MODEL TCL SECTION 4
WIRING
4.5 APPLY POWER TO THE ANALYZER AND COMPLETE QUICK START
1. Once all wiring connections are secured and verified, apply power to the analyzer.
2. When the analyzer is powered up for the first time, Quick Start screens appear. Using Quick Start is easy.
a. A backlit field shows the position of the cursor.
b. To move the cursor left or right, use the keys to the left or right of the ENTER key. To scroll up or down or to increase
or decrease the value of a digit, use the keys above and below the ENTER key. Use the left and right keys to move
the decimal point.
c. Press ENTER to store a setting. Press EXIT to leave without storing changes. Pressing EXIT also returns the
display to the initial Quick Start screen.
d. A vertical black bar with a downward pointing arrow on the right side of the screen means there are more items to
display. Continue scrolling down to display all the items. When you reach the bottom of the list, the arrow will point up.
FIGURE 4.4. Wiring Sensor with Optimum EMI/RFI
or Variopol Cable to Model 1056 Analyzer
FIGURE 4.5 Wiring Sensor with Standard Cable
to Model 1056 Analyzer
3. Choose the desired language. Scroll down to display more choices.
4. Choose total chlorine for sensor 1 (S1).
5. Choose the desired units for chlorine.
Language
Francais
Espanol
Deutsch
English
S1 Measurement
Free Chlorine
pH Independ. Free Cl
Total Chlorine
Monochloramine
Total Chlorine
Units
mg/L
ppm