PROGRAMMING THE ANALYZER
TABLE 7-1. DEFAULT SETTINGS
ITEM CHOICES DEFAULT
Outputs
1. Assignments
a. output 1
chlorine, temp chlorine
b. output 2
chlorine, temp temp
2. Range 0-20 or 4-20 mA 4 – 20 mA
3. 0 or 4 mA setting
a. chlorine
-9999 to +9999 0
b. temperature
-999.9 to +999.9 0
4. 20 mA setting
a. chlorine
-9999 to +9999 10
b. temperature
-999.9 to +999.9 0
5. Fault current (fixed) 0.00 to 22.00 mA 22.00 mA
6. Dampening
0 to 999 sec 0 sec
7. Simulate
0.00 to 22.00 mA 12.00 mA
Alarms
1. Logic
high or low AL1 low, AL2,3,4 high
2. Assignments
a. AL1 and AL2
chlorine, temp, fault,
interval timer
chlorine
b. AL3 and AL4
chlorine, temp, fault,
interval timer
temperature
3. Deadband 0 to 9999 0
4. Interval timer settings
a. interval time
0.0 to 999.9 hr 24.0 hr
b. on time 0 to 999 sec 10 sec
c. recovery time 0 to 999 sec 60 sec
Measurement (Chlorine)
1. units ppm or mg/L ppm
2. resolution 0.01 or 0.001 0.001
3. input filter 0 to 999 sec 5 sec