51
NOTE
Do not use a sharp probe. It will puncture the bladder and destroy the sensor.
Gently press the probe against the bladder several times to force liquid through the holes at the base of the
cathode stem. Keep pressing the bladder until no air bubbles can be seen leaving the holes. Be sure the holes
remain covered with electrolyte solution.
10. Place a drop of electrolyte solution on the cathode, then place the membrane assembly over the cathode.
Screw the membrane retainer in place.
11. The sensor may require several hours operating at the polarizing voltage to equilibrate after the electrolyte
solution has been replenished.
MODEL TCL SECTION 10
MAINTENANCE
33523-00 Electrolyte Fill Plug
9550094 O-Ring, Viton 2-014
33521-00 Membrane Retainer
23501-02 Total Chlorine Membrane Assembly: includes one membrane assembly and one O-ring
23502-02 Total Chlorine Membrane Kit: includes 3 membrane assemblies and 3 O-rings
9210438 Total Chlorine Sensor Fill Solution, 4 oz (120 mL)
TABLE 10-1. Spare Parts
FIGURE 10-2. Sensor Parts