– 3 –
For your safety
(General)
is the safety information.
CAUTION:
THIS PRODUCT UTILIZES A LASER.
USE OF CONTROLS OR ADJUSTMENTS OR PERFORMANCE OF PROCEDURES OTHER THAN
THOSE SPECIFIED HEREIN MAY RESULT IN HAZARDOUS RADIATION EXPOSURE. DO NOT OPEN
COVERS AND DO NOT REPAIR YOURSELF. REFER SERVICING TO QUALIFIED PERSONNEL.
NOTICE FOR LASER
CLASS 1
LASER PRODUCT
Inside of unit
NOTICE FOR MEDICAL USE
•
This is an apparatus intended for recording playing-back of diagnosis image classified as;
•
Protection against Electric Shock Class 1
•
Protection against Ingress of Water Ordinary
•
Mode of Operation Continuous
•
Not suitable for use in the presence of a Flammable Anaesthetic mixture with Air or with Oxygen or Nitrous Oxide.
•
Equipment connected to signal input and/or output parts must be certified according to the appropriate IEC 60601-1
and/or IEC 60601-1-1 harmonized national standard.
Furthermore all configurations shall comply with the system standard IEC 60601-1-1. Everybody who connects additional
equipment to the signal input part, or signal output part configures a medical system, and is therefore responsible that the
system complies with the requirements of the standard IEC 60601-1-1. If in doubt, consult the technical service department
or your local representative.
Leakage current of this unit may be exceeded the allowable value when conductively connected to other equipment.
To avoid increment of the leakage current, separation device shall be applied.
For your safety
(USA and Canada)