Reflection Standard
P/N 968-62
Serial No. CXXXXXXXP-XXXXX
IMPORTANT!
by smudge marks and dust; and must
be kept clean.
The standard is dramatically affected
Date XX/XX/XX
REFL
[%]
WAVE
[nm]
390
400
410
420
430
440
450
460
480
500
520
540
560
580
600
620
640
660
680
700
710
X-Rite
62.10
76.70
81.88
85.02
85.99
86.12
86.48
86.74
87.05
87.42
87.36
87.08
86.72
86.25
85.71
85.18
84.52
84.02
83.35
82.69
81.99
CALIBRATION
CHECK
C/2 (WHITE)
L*
a*
b*
Instrument Model No.
Instrument Serial No.
COLOR CHECK
Refer to section in operation manual
appendix for procedure to obtain
reference values.
DATE: TEMP:
ILLUM/OBS:
APERT:
REFERENCES VALUES
STEP
1 (White)
2 (Blue)
3 (Brown)
L* a* b*
Notes:
X-Rite Reflection Standard
The Calibration Values for the white spot are affected by the environment
and cleaning method of the standard.
If the white spot does not measure correctly, it could be that the unit
needs to be calibrated; there is dust in the optics; or the standard has
smudge marks or is dirty.
The ceramic standard should be cleaned using a mild soap and warm
water solution, thoroughly rinsed with warm water, and wiped dry with a
clean, lint free cloth.
If you lose your envelope, you can obtain the calibration values from the
back page of this manual, or, by looking at them in the calibration
procedure. Otherwise, you can contact X-Rite with the serial number of
your standard and get the values.
®