A SERVICE OF

logo

5. Setting System Parameters
5.1 Averaging
The averaging feature allows the unit to make several measurements from the same
and/or different locations on a sample. This will allow you to have a better overall
average of a sample color. The averaging feature has two functions:
SAMPLES AVERAGED - allows you to enter the number of measurements
you will make (1-99) at various locations on the sample, to obtain an average
value.
SUB AVERAGE - allows you to enter the number of measurements the unit
will take (1-6) on a sample at a single location, to obtain an average value.
> When average and sub average are set to one, no averaging is performed and no
averaging messages will be displayed.
1) Press both [FUNCTION] and [ILLUM]
at the same time.
> To exit Average Setup: Press
[FUNCTION] and [ILLUM] at the
same time and the unit will return
back to normal operation.
2) Select "NO", press [FUNCTION].
FUNCTION ILLUM
DIF
CAL
PRESS AT SAME TIME
X 4 . 1 2
Y 2 . 8 7 Z 2 . 4 1
C
2
X Y Z
FUNCTION ILLUM
DIF
CAL
CALIBRATE ?
YES
PRESS
NO
35