A SERVICE OF

logo

5.3 RS232 I/O Parameters
Your instrument comes equipped with a serial port that allows data to be
transmitted/received to/from an external device. Listed below are the available I/O
parameters.
"RCI ON/OFF" (Remote Control Interface) - enables or disables the ability
to externally control the 948/968 via the I/O port.
"BAUD" - determines the input/output rate (characters per second) of the
I/O port. Available outputs are: 300, 600, 1200, 2400, 4800, and 9600.
"CR WITH LF/CR WITH NO LF" (Carriage Return with Line
Feed/Carriage Return with no Line Feed) - varies the delimiter at the end of
each line of data. When set to CR WITH NO LINE FEED, just a Carriage
Return is sent at the end of a line of data. When set to CR WITH LF, a
Carriage Return and then a Line Feed are sent at the end of a line of data.
"HANDSHAKE" is used for data transmission. Handshake may be set to
HANDSHAKE OFF (ASCII = 13H), PIN 5 CTS (Clear to Send), PIN 5
BUSY, or XON/XOFF ENABLED (ASCII = 11H). Handshake should
normally be set to OFF when it is not being used.
1) Press both
[FUNCTION]
and
[ILLUM]
at
the same time.
> To exit I/O Setup: Press
[FUNCTION] and [ILLUM] at the
same time and the unit will return
back to normal operation.
2) Select "NO", press
[FUNCTION]
three
times.
FUNCTION ILLUM
DIF
CAL
PRESS AT SAME TIME
X 4 . 1 2
Y 2 . 8 7
Z 2 . 4 1
C
2
X Y Z
FUNCTION ILLUM
DIF
CAL
CALIBRATE ?
YES
PRESS
NO
THREE
TIMES
39