Eizo RS110 Flat Panel Television User Manual


 
Recommended separation distances between portable and mobile RF communications
equipment and the RadiForce Series
The RadiForce series is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the RadiForce series can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the RadiForce series as recommended below, according to
the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
Separation distance according to frequency of transmitter
m
150kHz to 80MHz
d = 1.2
P
80MHz to 800MHz
d = 1.2
P
800MHz to 2.5GHz
d = 2.3
P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance
“d” in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where
“P” is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Cable length
Power Cord : 3.0m
Signal Cable (DD300) : 3.0m
Signal Cable (MD-C87) : 1.8m
USB Cable (UU300) : 3.0m