VXI SVM2608 TV Converter Box User Manual


 
VXI Technology, Inc.
44 SVM2608 Programming
Bit 14 - Measures 128 on the 10 k scale
Bit 15 - Measures 81.92 k on the 100 k scale
Bit 16 - Measures 81.92 k on the 1 M scale
Bits 17 – 31 are not used and read as “0”.
For Channels 4-5:
Bit 0 - Measures +0.117 V on the 0.5 V scale
Bit 1 - Measures -0.117 V on the 0.5 V scale
Bit 2 - Measures +0.945 V on the 1 V scale
Bit 3 - Measures -0.945 V on the 1 V scale
Bit 4 - Measures +0.945 V on the 2 V scale
Bit 5 - Measures -0.945 V on the 2 V scale
Bit 6 - Measures +0.945 V on the 5 V scale
Bit 7 - Measures -0.945 V on the 5 V scale
Bit 8 - Measures +9.45 V on the 10 V scale
Bit 9 - Measures -9.45 V on the 10 V scale
Bit 10 - Measures +9.45 V on the 20 V scale
Bit 11 - Measures -9.45 V on the 20 V scale
Bit 12 – 31 are not used and read as “0”.
Example 1
Reading 0x0000000C after a Self Test command indicated a problem during Self Test 0 and Self
Test 1, measuring ±0.945 V on the 2 V scale, but no problem on all other tests. This would
indicate that the front end programmable gain amplifier has a problem on the x5 (gain = 5) setting.
Example 2
Reading 0x00000000 after a Self Test operation would indicate that all Self Tests passed
successfully.
Preset Setting Measurement Commands
The following measurement commands are executed by the microprocessor using pre-set, factory-
programmed settings:
0x0005 = 100 Range Query (2-wire)
0x0006 = 1 k Range Query (2-wire)
0x0007 = 10 k Range Query (2-wire)
0x0008 = 100 k Range Query (2-wire)
0x0009 = 1 M Range Query (2-wire)
0x000A = Auto-range Query (2-wire)
0x000B = 100 range query (4-wire)
0x000C = 1 k range query (4-wire)
0x000D = 10 k range query (4-wire)
0x000E = 100 k range query (4-wire)
0x000F = 1 M range query (4-wire)
0x0010 = Auto-range query (4-wire)
0x0012 = 1 V Measurement
0x0013 = 2 V Measurement
0x0014 = 5 V Measurement
0x0015 = 10 V Measurement