Procedure
See Figure 2-2 for information on installing the meter (with a weldolet fitting) in a free-
stream application.
Important
You must always install the meter horizontally and oriented to allow flow in the gap between the
tines, irrespective of the pipeline orientation (horizontal or vertical). This position helps to prevent
the trapping of bubbles or solids on the meter – allowing the solids to sink and the bubbles to rise.
Free-stream (weldolet fitting) meter installationFigure 2-2:
STATUS
SCROLL SELECT
A
C
D
B
Plan view of a vertical pipe installation
A. 4-inch pipe for horizontal installations; 6-inch (152 mm) pipe for vertical installations
B. 2.1 in (52.5 mm) meter opening in pipeline
C. Weld
D. Free-stream weldolet (purchased to fit pipe diameter)
2.3 Mount with a T-piece (flanged fitting)
Prerequisites
T-piece (flanged) installations are recommended for processes with the following
conditions:
Flow 0.3 to 0.5 m/s (at the pipe wall)
Viscosity 0.5 to 100 cP
Temperature - -50 °C to 200 °C (–58 °F to 392 °F)
- -40 °C to 200 °C (-40 °F to 392 °F) in
hazardous areas
Mounting