1 Planning
Topics covered in this chapter:
Installation checklist
Best practices
Power requirements
Other installation considerations
Recommended installations for short-stem meters
Perform a meter check (pre-installation)
1.1 Installation checklist
Verify the contents of the product shipment to confirm you have all parts and
information necessary for the installation.
Verify the meter calibration range and boundary corresponds to the planned
installation. A calibration mismatch can cause measurement error, and will need to
be corrected.
Make sure that all electrical safety requirements are met for the environment in
which the meter will be installed.
Make sure that the local ambient and process temperatures and process pressure
are within the limits of the meter.
Make sure that the hazardous area specified on the approval tag is suitable for the
environment in which the meter will be installed.
Make sure that you will have adequate access to the meter for verification and
maintenance.
Verify that you have all equipment necessary for your installation. Depending on
your application, you may be required to install additional parts for optimal
performance of the meter.
If your meter will be wired to a remote-mount 2700 FOUNDATION fieldbus
transmitter:
- Refer to the instructions in this manual for preparing the 4-wire cable and wiring
to the processor connections.
- Refer to the instructions in the transmitter installation manual for mounting and
wiring the 2700 FOUNDATION fieldbus
transmitter. See Micro Motion
Model 1700 and Model 2700 Transmitters: Installation Manual.
- Consider the maximum cable length between the meter and transmitter. The
maximum recommended distance between the two devices is 1000 ft (300 m).
Micro Motion recommends using Micro Motion cable.
Planning
Installation Manual 1