Emerson Micro Motion Fork Viscosity Meters TV Antenna User Manual


 
Power cable recommendations for explosion-proof/flameproof meters
Minimum wire gauge (AWG per feet)Figure 1-1:
300 600 900 1200 1500 1800 2100 2400 2700 3000
Distance of Installation (ft)
21.6V
24V
14
15
16
17
18
19
20
21
22
23
24
25
26
AWG Maximum
Minimum Wire Gauge
Minimum wire area (mm
2
per meter)Figure 1-2:
0.000
0.050
0.100
0.150
0.200
0.250
0.300
0.350
0.400
100 200 300 400 500 600 700 800 900 1000
Minimum Wire Area (mm
2
)
Distance of Installation (m)
Minimum Wire Area (mm
2
)
21.6V
24V
Planning
Installation Manual 3