Fluke VR101S Cable Box User Manual


 
VR101S
Users Manual
54
Sag, Swell & Outage Measurement
Hot to Neutral
Model number VR101/xx3
Range:
0 to 200Vrms
Accuracy:
±2Vrms*
Resolution:
1Vrms
Model numbers VR101/xx1, /xx4, /xx5, and /xx7
Range:
0 to 270Vrms
Accuracy:
±4Vrms**
Resolution:
2Vrms
Neutral to ground
Model number VR101/xx3
Range:
3 to 150Vrms
Accuracy:
±2Vrms
Resolution:
1Vrms
Model numbers VR101/xx1, /xx4, /xx5, and /xx7
Range:
3 to 150Vrms
Accuracy:
±2Vrms
Resolution:
1Vrms
* sags less than 70V for > 1s will be reported as 0V
** sags less than 140V for > 1s will be reported as 0V