1.4 SPECIFICATIONS — MODEL 499ACL-02 SENSOR
1.5 PERFORMANCE SPECIFICATIONS — COMPLETE SYSTEM
Wetted Parts: Gold, Noryl
®1
(PPO), Viton
®2
, EPDM, Silicone
Dimensions: 1.0 x 5.6 in. (25.4 x 143 mm)
Cable: 25 ft. (7.6m) standard
Pressure Rating: 0 to 65 psig (101 to 549 kPa)
Temperature Rating: 32 to 122
°
F (0 to 50
°
C)
Electrolyte Capacity: Approximately 25 mL
Electrolyte Life: Approximately 4 months
Weight/Shipping Weight: 1 lb/3 lb (0.5 kg/1.5 kg)
1
Noryl is a registered trademark of General Electric.
2
Viton is a registered trademark of DuPont Performance Elastomers.
Linear Range: 0 to 20 ppm (mg/L) as Cl
2
(for higher ranges, consult factory)
Linearity (per ISO 15839): 0-10 ppm: 2%; 0-20 ppm: 3%
Response Time: Following a step change in concentration, the reading reaches 90% of final value within 7 minutes at
25°C.
Drift: At about 1.5 ppm in clean water and constant temperature, drift is typically less 0.05 ppm over two weeks.
Detection Limit (per ISO 15839): 0.02 ppm (mg/L) in clean water at room temperature