Emerson Process Management TLC TV Converter Box User Manual


 
MODEL TCL SECTION 7
SOFTWARE CONFIGURATION
ITEM CHOICES FACTORY SETTINGS
CONFIGURE (continued)
C. Alarms (Section 7.7) (continued)
4. Alarm 2 Control
a. Activation method Chlorine, temperature Chlorine
b. Control mode Normal or TPC Normal
5. Alarm 2 setup (normal)
a. Configuration Low, high, or off Low
Rest of alarm 2 setup is the same as alarm 1
6. Alarm 3 control and setup is the same as alarm 1
7. Alarm 4 control
Alarm Fault or off Fault
8. Feed limit timer
a. Feed limit Disable, alarm 1, alarm 2, or alarm 3 Disable
b. Timeout value 0 to 10,800 sec 600 sec
9. Interval timer
a. Select alarm Disable, alarm 1, alarm 2, or alarm 3 Disable
b. Interval time 0 to 999.9 hr 24.0 hr
c. Repeats 1 to 60 1
d. On time 0 to 2999 sec 120 sec
e. Off time 0 to 2999 sec 1 sec
f. Recovery time 0 to 999 sec 600 sec
D. Temperature compensation (Section 7.9)
1. Temperature compensation Auto or manual Auto
2. Manual temperature -15 to 130°C 25°C
E. Noise Reduction (section 7.10)
Noise rejection 50 or 60 Hz 60 Hz
F. Main sensor calibration (Section 7.11)
1. Stabilize reading 0 to 9999 ppm 0.05 ppm
2. Stabilize time 0 - 30 sec 10 sec
3. Sensor zero stabilization value
4. Dual range calibration Enable or disable disable
G. Security (Section 7.13)
1. Lock all 000-999 (000 disables) 000
2. Lock program 000-999 (000 disables) 000
3. Lock configuration 000-999 (000 disables) 000
TABLE 7-1. Program Settings List (continued)
21