Emerson Process Management TLC TV Converter Box User Manual


 
20
MODEL TCL SECTION 7
SOFTWARE CONFIGURATION
Continued on the following page
ITEM CHOICES FACTORY SETTINGS
CONFIGURE
B Outputs (Section 7.6)
1. Output 1 Control
a. Measurement Oxygen, chlorine, ozone, pH, or temperature Oxygen
b. Control Normal or PID Normal
2. Output 1 Setup (normal)
a. Current 4-20 mA or 0-20 mA 4-20 mA
b. Dampening 0-299 sec 0 sec
c. Hold mode Hold last value or go to fixed value Hold last value
d. Fixed hold value 0-22 mA 21 mA
e. Fault value 0-22 mA 22 mA
3. Output 1 Setup (PID)
a. Proportional 0 to 299.9% 100 %
b. Integral 0 to 2999 sec 0 sec
c. Derivative 0 to 299.9% 0/0%
4. Output 2 Control
a. Measurement Oxygen, chlorine, ozone, pH, or temperature Temperature
b. Control Normal or PID Normal
5. Output 2 Setup (normal) See output 1 See output 1
6. Output 2 Setup (PID) See output 1 See output 1
7. Hold feature Enable, disable, or 20 min timeout Disable
C. Alarms (Section 7.7)
1. Alarm 1 Control
a. Activation method Chlorine or temperature Chlorine
b. Control mode Normal or TPC Normal
2. Alarm 1 setup (normal)
a. Configuration Low, high, or off High
b. Hysteresis
if chlorine -9999 to 9999 ppm 0 ppm
if temperature 0 to 10°C 0.1°C
c. Delay time 0-99 sec 0 sec
d. Relay fault none, open, closed None
3. Alarm 1 setup (TPC)
a. Setpoint
if chlorine -9999 to 9999 ppm 1.0 ppm
if temperature -5 to 130°C 25°C
b. Proportional 0 to 299.9% 100 %
c. Integral 0 to 2999 sec 0 sec
d. Derivative 0 to 299.9% 0.0%
e. Time period 10 to 2999 sec 30 sec
f. LRV (100% on)
if chlorine -9999 to 9999 ppm 0 ppm
if temperature -5 to 130°C 0°C
g. URV (100% off)
if chlorine -9999 to 9999 ppm 2 ppm
if temperature -5 to 130°C 100°C
h. Relay fault None, open, or closed None
TABLE 7-1. Program Settings List (continued)