Emerson Process Management TLC TV Converter Box User Manual


 
46
MODEL TCL SECTION 9
CALIBRATION - TOTAL CHLORINE
SECTION 9
CALIBRATION - TOTAL CHLORINE
9.1 INTRODUCTION
The continuous determination of total chlorine requires two steps. See Figure 9-1. First, the sample flows into a
conditioning system (Model TCL) where it is treated with acetic acid (vinegar) and potassium iodide. The acid low-
ers the pH, which allows total chlorine in the sample to quantitatively oxidize the iodide to iodine. The treated sam-
ple then flows to the sensor. The sensor is a membrane-covered amperometric sensor, whose output is propor-
tional to the concentration of iodine. Because the concentration of iodine is proportional to the concentration of
total chlorine, the analyzer can be calibrated to read total chlorine.
Figure 9-2 shows a typical calibration curve for a total chlorine sensor. Because the sensor really measures iodine,
calibrating the sensor requires exposing it to a solution containing no iodine (zero standard) and to a solution con-
taining a known amount of iodine (full-scale standard).
The zero standard is necessary because the sensor, even when no iodine is present, generates a small current
called the residual current. The analyzer compensates for the residual current by subtracting it from the measured
current before converting the result to a total chlorine value. New sensors require zeroing before being placed in
service, and sensors should be zeroed whenever the electrolyte solution is replaced. Deionized water is a good
zero standard.
The purpose of the full-scale standard is to establish the slope of the calibration curve. Because stable total chlo-
rine standards do not exist, the sensor must be
calibrated against a test run on a grab sample
of the process liquid. Several manufacturers
offer portable test kits for this purpose.
Observe the following precautions when taking
and testing the grab sample.
Take the grab sample from a point as close
as possible to the inlet of the TCL sample
conditioning system.
Total chlorine solutions are unstable. Run
the test immediately after taking the sam-
ple. Try to calibrate the sensor when the
chlorine concentration is at the upper end
of the normal operating range.
The Model 499ACL-02 (total chlorine) sensor
loses sensitivity at high concentrations of chlo-
rine. The 54eA controller has a dual slope fea-
ture that allows the user to compensate for the
non-linearity of the sensor. However, for the
vast majority of applications, dual slope cali-
bration is unnecessary.
FIGURE 9-1. Determination of Total Chlorine
FIGURE 9-2. Sensor Current as a Function of Total
Chlorine Concentration