7.1 CHANGING ALARM SETPOINTS
2. Press any key to enter the main menu. Move the cursor to "Program"
and press Enter (F4).
3. Press Enter (F4).
4. Move the cursor to the desired alarm and press Enter (F4).
5. The screen appearing at this point depends on how the alarm was
configured.
6. If the alarm is a normal (i.e., not TPC) alarm, a screen like the one
shown will appear. The alarm is a low alarm and the setpoint is 0.00
ppm. Press Edit (F4). Use the arrow keys to change the setpoint.
Press Save (F4) to store the new value. Press Exit (F1) to return to
the screen in step 4. Choose a new alarm.
7. If the alarm is TPC, a screen like the one shown will appear. The set-
point is +1.000 ppm. Press Edit (F4). Use the arrow keys to change
the setpoint. Press Save (F4) to store the new value. Press Exit (F1)
to return to the screen in step 4. Choose a new alarm.
Alarm setpoints
Output setpoints
Simulate tests
Exit Enter
Alarm 1 setpoint
Alarm 2 setpoint
Alarm 3 setpoint
Exit Enter
Alarm Low : 0.000 ppm
Exit Edit
Setpoint : 1.000 ppm
Exit Edit
1. Before changing alarm setpoints, be sure that alarms are properly configured. See Section 7.7.
MODEL TCL SECTION 7
SOFTWARE CONFIGURATION
25