MODEL TCL SECTION 7
SOFTWARE CONFIGURATION
7.5 CHOOSING DISPLAY OPTIONS
3. Press any key to enter the main menu. Move the cursor to "Program"
and press Enter (F4).
4. Move the cursor to "Configure" and press Enter (F4).
5. With the cursor on "Display", press Enter (F4).
6. A screen showing the present main measurement will appear. To
change the measurement to total chlorine, press Edit (F4), then use
the key to scroll through the choices until total chlorine is high-
lighted. Press Save (F4) to store the setting.
A screen will appear warning that if the measurement is changed, the
analyzer will return to factory default settings. Press Cont (F3) to con-
tinue. Press Abort (F1) to cancel the change.
continued on following page
Display
Outputs
Alarms
Exit Enter
1. The 54eA controller can be used with any amperometric sensor manufactured by Rosemount Analytical. The
default sensor is oxygen. In order to use the controller with a total chlorine sensor, the sensor identification
must be changed.
2. The display menu also lets the user customize the third line in the display, change timeout values, choose a
language other than English, and change the display contrast.