15
Type in threshold value; press
Note: When selecting a threshold torque value the ETA will ignore any
torque that is below this value. You cannot take another reading until
the applied torque falls below this value.
Type in amount of Dip; press
Select Second Parameter using ; press
If the setup and transducer has not been changed since the ETA was last used you are now
ready to take torque measurements.
Pressing after taking a series of readings will result in statistical information being dis-
played (Range, mean and standard deviation) for the stored readings.
To return to measure mode, press
If you have selected ‘Click Dip’ type in the amount of ‘Dip’ to be sensed; press
Note: 5% of torque for 0 to 100 Nm and 10% of torque for over 100 Nm
generally provide excellent results.
You are now ready to take torque measurements. When finished press
Note: When taking readings you can change the set-up parameters by
pressing the up arrow
Changing any of the set parameters will result in the ETA prompting to erase all data.
To erase stored readings, press
To return to setup screens, press
Example: You have been measuring in a right hand direction and you want to
switch to left hand.
Press the up arrow twice, change from right hand to left hand. Screen will prompt to erase
all data.
Press to erase all stored readings and return to the measure screen.
Enter
Enter
Enter
#
#
Enter
MENU
Enter
MENU
Enter
7529_eng.fm Page 15 Wednesday, November 12, 2003 8:57 AM