3. Technical Data
according to IEC-1010; UL 3101
Type
Nominal voltage (±10 %)
Nominal frequency
Power draw
Max. start-up current for 5 sec.
Protective class
Circuit breaker:
(MDA made by Schurter)
Pollution degree
Overvoltage installation category
Heat emission (max.)
Approval
-1 -2 -3
100 VAC 120 VAC 230 VAC
50/60 Hz 60 Hz 50 Hz
1900 VA 1900 VA 1900 VA
35A eff. 35 A eff. 25 A eff.
III
15A M3 15A M3 15A M3
222
II II II
1900 J/s 1900 J/s 1900 J/s
CE/c_CSA_us CE/c_CSA_us CE/c_CSA_us
All specifications related to temperature are valid only up to an ambient temperature of 18°C
to 35°C and a relative humidity of no more than 60%.
Microtome
Type: Rotary microtome,
encapsulated
Section thickness range 1 to 100μm
Trimming range:
Clinic 10 - 40μm
Research 1 - 600μm
Total specimen feed 25mm + 1mm
Vertical stroke 59mm ±0.5mm
Specimen retraction 20μm (can be deactivated)
Maximum specimen size 50 x 80mm
Cutting speed Slow: 0-50 strokes/min
Fast: 0-85 strokes/min
Vmax: 85-90 strokes/min
Specimen orientation 8° (x, y-axis)
Electric coarse feed
Slow 300μm/s
Fast 900μm/s
Cryostat
Dimensions:
Width (w/o handwheel) 700 mm
Width (with handwheel) 835 mm
Depth (cabinet only) 850 mm
Height (total) 1215 mm
Working height (armrest) 1025 mm
Weight:
Weight (w/motor and suct.) 193 kg
Weight (w/motor, w/o suct.) 185 kg
Weight (w/o motor, w/ suct.) 183 kg
Weight (w/o motor, w/o suct.) 175 kg
Weight (w/ specimen head cool.) 165 kg
Weight (w/o specimen head cool.) 145 kg
For "Site Requirements," refer to item 6.1
(page 20).
Lamp
50/60 Hz-Version: Osram DULUX L 18W/840
General information
Operating temperature range: 18°C to 35°C
Temperature range
during storage: +5°C to +55°C
Relative humidity: max. 60%,
non-condensing
Storage humidity: < 60%