6. Installation
6.1 Site Requirements
The place of installation must meet the following requirements:
• No direct sunlight.
• Power supply at a distance no greater than approx. 3 m.
• No drafts (air condition outlets etc.).
• Even floor.
• Vibration-free floor.
• The instrument is designed for indoor use only.
• Obstruction-free access to the handwheel.
• The power switch/circuit breaker must be freely and easily accessible.
• Room temperature always approx. 22°C.
• The relative humidity must not exceed 60%.
To ensure proper function of the instrument, it must be set up while
maintaining a minimum distance from walls and furniture.
• Distance to walls and furniture, calculated from the cabinet:
- rear: 15 cm
- right side: 30 cm
- left side: 15 cm
• No heat dissipating appliances around.
High room temperatures and excessive air humidity affect the
cooling capacity of the cryostat and lead to ice forming in the
instrument!
• Plug the instrument into power sockets with ground only.
Power cable length: up to 3.5 m
Extendable: no