A few rules:
• Always begin at a low extraction level, then slowly increase it.
• Do not use high extraction levels unless absolutely necessary.
• Different specimen sizes require different extraction levels.
• The faster the trimming or sectioning speed, the lower the extraction
level should be.
• The larger and/or thicker the specimen to be trimmed, the lower the
extraction value.
• When section specimens with a diameter of 0.5 cm, the anti-roll guide
stretches the section adequately. For larger specimens, we recommend
using the vacuum function.
8. Working with the Instrument
Trimming, 1 x 1cm specimen
Trimming thickness Speed Extract. value
10μm 40% 1.5
10μm 60% 1.5
10μm 80% 1
10μm 100% 1
20μm 40% 2
20μm 60% 2
20μm 80% 1.5
20μm 100% 1
30μm 40% 2.5
30μm 60% 2.5
30μm 80% 2
30μm 100% 1.5
40μm 40% 3
40μm 60% 3
40μm 80% 2.5
40μm 100% 2
Trimming, 1 x 1cm specimen
Trimming thickness Speed Extract. value
1μm - 5 μm 40% 2.5
60% 2
80% 1.5
100% 1.5
10μm 40% 2.5
60% 2
80% 2
100% 1.5
The values listed here are possible
settings that depend on the tissue
type, specimen size, section thick-
ness and section speed, and are in-
tended for orientation purposes only.
These settings can also vary when using
different anti-roll plate!