Leica CM1950 – Cryostat
Installation of the new lamp
Type:
OSRAM DULUX L 18W/840.
• Hold the lamp in the correct mounting posi-
tion as shown and push it to the left until it
engages in the holder.
• Slightly lift the sliding window (1) holding it by
the grip (2) and pull it out to the front.
1
2
34
Removal of the lamp
• The lamp is mounted behind a glare shield and
therefore not visible.
• Touch the lamp for better orientation.
• Lightly tilt the fluorescent tube down to the left
and pull it out of the clip (3).
• Hold the lamp with both hands and pull it to
the left out of the holder (4).
Only use lamps of the same specifica-
tion!
Fig. 70
Fig. 69
Fig. 68