Leica CM1950 Projection Television User Manual


 
55
Leica CM1950 – Cryostat
Assembling the anti-roll guide system
(for the blade holder CE)
1. Insert the glare-minimizing glass insert into the
interchangeable frame and tighten it evenly
using the knurled screw (10).
2. Insert the shaft (11) of the metal frame for
exchangeable glass inserts from above into
the bore of the swinging arm (12) in such a way
that the pin rests in the notch.
3. Push the white plastic plate (14) from below
onto the shaft (11).
4. Screw the knurled nut (13) from below onto
the shaft (11).
The glass of the anti-roll guide can be
used from all 4 sides when replace-
ment is necessary (the glass stage
plate can be reordered).
13
14
11
6
10
10
With blade ejector (1)
With knife guard (2)
Integrated lateral shift and stable base
Clearance angle adjustment (5) with 3 mm
Allen key (see detail image 5, to left at knife
holder) - recommended angle from 2° to 5°.
With anti-roll guide (4)
Lever (6) for the lateral shift
Lever (7) for clamping the blade
Lever (8) for clamping the base (9) to the
dovetail guide in the chamber
Pressure plate (3) for section extraction
When using low-profile blades, the ruler
(7, Fig. ) must be inserted.
Blade holder CE with anti-roll guide for low-profile (NB) and high-profile (BB)
1
4
5
2
3
11. Optional Accessories
Fig. 41
Fig. 41a
Fig. 42
4
6
7
8
9
12