Leica CM1950 Projection Television User Manual


 
56
Operating Manual V1.1 – 05/2007
The knife guard is folded
away in the detail view.
Knife guard folded away
Fig. 43
Four possible
settings
Blade rest (ruler) for low-profile
blades
1
2
1
2
6
Fig. 44
11. Optional Accessories
Red elements on the blade or knife
holders, such as the knife guard and
ejector, are protective devices that
may not be removed.
3
4
7
The anti-roll guide and
pressure plate must be
parallel to each other.
6
Blade ejector (1) and knife guard (2)
Integrated lateral shift and stable base
Clearance angle adjustment using size 3 Allen
key (recommended angle between 2° and 5°)
With finger rest (6) for brush specimen
Clamping lever (8) for lateral shift must point
downward to permit unhindered shifting of
finger rest.
when using BB knives, remove the knife
support (7).
Conversion of blade holder with anti-roll guide
to blade holder with finger rest
Screw off the anti-roll guide.
Unscrew left Allen screw using 2.5 mm Allen
key and remove base of anti-roll guide.
Attach the finger rest (6) from the left, tighten
the hexagon-head screw using the No. 2.5
key – be careful of the blade ejector!
If you are working with the brush
technique, the knife guard must be
folded upwards.
Blade holder CE with finger rest (for NB + BB)
7
8
5