Leica CM1950 Projection Television User Manual


 
61
Leica CM1950 – Cryostat
Knife holder CNZ with anti-roll guide
Pressure plate permits the full length of the
knife to be utilized.
Use hard metal and steel knives here.
Fig. 52
11. Optional Accessories
The height of resharpened knives must be adjusted using knurled
screws (3) (approx. 1mm under the edge of the clamping jaw).
Ensure that the knife is adjusted in parallel vertically.
Knife guard/lateral movement for knife
holder CN
The knife guard (18) is fixed and integrated into
the clamping jaws. The knife guard has handles
(19) that allow it to be moved. The knife guard is
adequate for knives up to 22 cm in length. Always
cover exposed parts of the knife blade after
sectioning.
The anti-roll guide system can be moved side-
ways (only for the 84 mm variant). To better find
the mid position, a groove (17) is provided in the
shaft (16).
19
19
18
16
17
Fig. 53