Agilent Technologies 8922 VCR User Manual


 
12-34
Module I/O Specifications
A16 Receiver
To test FM Demod apply RF Carrier with FM modulation to RF Input, measure FM
Demod on oscilloscope.
NOTE: Ensure correct settings on RF Analyzer page (frequency/amplitude). If the RF Input level
is greater than 5 dBm below RF Analyzer setting, the measurement will not register.
Refer to "A9 Global Test and Demod", page 12-15, for measurement procedure.
To Front Panel
FM_DEMOD J1(13)
Sensitivity into open circuit: 20 µV/Hz
Output Impedance: 600
Max Deviation: 100 kHz peak
Accuracy DC to 270 kHz: ± 5%
Sensitivity into oscilloscope 325 µV/Hz
To A3 Audio Analyzer
DEMOD_AUD J2(6)
Output impedance: <10
Maximum Voltage Output: 12 Vp
DC coupled AM
Depth: 0 to 95%
Sensitivity: 0.01 V/% AM
DC coupled FM
Max Deviation: 100 kHz
Sensitivity: 20 µV/Hz
To A9 Global Test and Demod
10.7M_IF J3(9)
Freq: 10.7 MHz ± 50 kHz
Level: +3 dBm ± 0.2 dB
Output Impedance: 50
Global Phase Error: < 0.8° RMS
< 1.5° Peak
Global Freq Error: < 1.5 Hz
Harmonics: < 40 dBc