Agilent Technologies 8922 VCR User Manual


 
3-3
Verifying Performance
Using the Compatibility Switch for the HP/Agilent 8922F/H or M/S
To Load the Program in the Agilent 8922M/S.
To verify the performance of the HP/Agilent 8922H/M you need to convert the instrument
back from an HP/Agilent 8922G, or convert the HP/Agilent 8922F/S to an HP/Agilent
8922E.
You are now ready to run the Performance Test Software.
1) Put the disk in the disk drive.
2) Type ``LOAD "PT_8922"'', press ENTER.
After you have completed the Performance Tests, return the instrument back to the
HP/Agilent 8922F/S or HP/Agilent 8922H/M using the same process in reverse.
Using the Compatibility Switch for the
HP/Agilent 8922F/H or M/S
Back Conversion
To turn the instrument from the HP/Agilent 8922H/M or HP/Agilent 8922F/S back to an
HP/Agilent 8922G or an HP/Agilent 8922E, select the following keys:
CONFIG (this is accessible from the Cell Control screen in the bottom right-hand
corner).
Compatible, select HP 8922G or HP 8922E
HP-IB Adrs (22)
PRESET
The instrument is now set up as an HP/Agilent 8922G or HP/Agilent 8922E and ready for
Performance Verification testing.
Forward Conversion
To return the instrument from an HP/Agilent 8922G back to an HP/Agilent 8922H/M or
an HP/Agilent 8922E to an HP/Agilent 8922F/S, select the following keys:
More (this is accessible from the Cell Control screen in the bottom right-hand corner).
Scroll down the list and select CONFIG.
Compatible, select HP 8922H/M or HP 8922F/S
HP-IB Adrs (14)
PRESET
The instrument is returned to an HP/Agilent 8922H/M or HP/Agilent 8922F/S.