Anritsu MP1777A Universal Remote User Manual


 
Appendix
A
SCPl
Error Messages
A.2
Run
Time
Errors
The error codes, [-299 and -2001, indicate that errors have taken place in the execution controller.
In this case, bit
4
of the event status register is set.
Errors take place when the following events take place
When the <PROGRAM DATA> to follow the header comes outside the specified range.
The program message cannot be executed because of the device status.
Table
A-2
Error Messages of
Run
Time Errors
Code Message Error Detection Condition
-220 Parameter error An error was found in the parameter.
-22
1
Setting conflict
Although the setting is correct as a parameter, it cannot be set because of the
device status.
Example:
A
self-diagnosis was attempted during measurement.
A rate greater than that of the interface was specified as the
final MUX stage.
-222 Data out of range
The numeric data has exceeded the device specifications.
Example:
SENSe
:
MEASure
:
PERiod
0,
s
:
SYSTem
:
MEMory
:
STORe 25
-223 Too much data
The
length of the string data has exceeded the specified length.
Example:
The following setting was used when word data of 16 characters
was specified.
:
SENSe
:
TELecom
:
PATTern
:
UWORd
"
1
1001 1001 1001 1001
"
-224 Illegal parameter value Using the parameter received is disabled.
Example:
1.
The character data comes outside the specified range.
:
SENSe
:
TELecom
:
BRATe M 15
2. The string data comes outside the specified range.
:
CALCulate
:
DATA'? "LAST
:
EC
:
CMI"
-240 Hardware error
The command cannot be executed because of hardware failure
-241 Hardware missing
The command cannot be executed because the option is not mounted.