PSC Power On Status Clear Command
Parameter
Function
<DECIMAL NUMERIC PROGRAM DATA>
O
...................
Sets the power ON status clear flag to false.
1
...................
Sets the power ON status clear flag to true.
Determines whether or not to clear each enable register of the service request, standard event
status and parallel poll of the status report module when the power is switched
ON.
When the mode is set to
0.
the enable register will not be cleared and the device is enabled to
generate SRQ after the power source is turned ON.
When the mode is set to
1,
the enable register will be cleared and the device is disabled to
generate SRQ after the power source is turned ON.
Example of use
Generates SRQ without clearing the power ON status flag.
>
vPSC
0
;
*SRE
32
;
*ESE
128
*PSC?
Power
On
Status
Clear
Query
Response
<NR1 NUMERIC RESPONSE DATA>
0
...................
Sets the power ON status clear flag to false.
1
...................
Sets the power ON status clear flag to true.
Function
Inquires about the truth or falsity of the power
ON
status clear flag.
Example of use
2
*PSC?
<
0
*
SAV
Save
Command
Parameter
Function
<DECIMAL NUMERIC PROGRAM DATA>
1
to 10
.........
4-4
Writes the current device setting into the memory of specified number.
There is no limit to the state to be saved. See the Table
7-
1
in the Section
7.5
"Parameter".
As an SCPI command that has the same function,
:
SYSTem
:
MEMory
:
STORe is avail-
able.
Example of use
Writes the current setting into the memory number
1.
z
-SAV
1