Anritsu MP1777A Universal Remote User Manual


 
Section
5
Common
IEEE488.2
Commands
*CLS Clear Status Command
Parameter None
Function
Clears all the status composition excluding the output queue and MAV summary message
However, the enable register and Transition filter are not cleared.
Clears both the output queue and MAV bit when
*CLS
is
sent immediately after
the
<PRO-
GRAM MESSAGE TERMINATOR>, and at the same time, before the <Query MESSAGE
UNIT> element.
Example of use
>
*CLS
-ESE
Standard Event Status Enable Command
Parameter <DECIMAL NUMERIC PROGRAM DATA>
Integer value between
0
and
255.
Set as the parameter the sum total of the bits desired to be enabled among the standard event
status enable register.
Set
1
for enable and
0
for disable.
See the Section for the "Status Byte" for the composition of the register of this device.
Function
Sets and clears the standard event status enable register.
Example of use
Sets bits
2
and
4
of the standard event status enable register.
>
*ESE
20
*ESR?
Standard Event Status Register Query
Response
Function
Example of use
<NU
1
NUMERIC RESPONSE DATA>
Integer value between
0
and
255.
Set
as
the response the sum total of the bita of the standard event status register.
See the Section for the "Status Byte" for the composition of the register of this device.
Inquires about the current value
of
the standard event status registel
When a command error is found.
>
*ESR?
<
32