Device Message Details
:
SOURce
:
TELecom
:
OFFSet <numeric>
Command type Sequential command
Parameter
<numeric>
=
<DECIMAL NUMERIC PROGRAM DATA>
-50 to
+50
...
step
0.
I
Function
Performs the frequency offset setting.
Example of use
Sets the value of frequency offset at +25.
>
:
SOURce
:
TELecom
:
OFFSet 25
:
SOURce
:
TELecom
:
OFFSet?
Command type Sequential command
Response <NR
1
NUMERIC RESPONSE DATA>
-50 to +50
...
step 0.1
Function
Makes a query of the value of frequency offset.
Example of use
>
:
SOURce
:
TELecorn
:
OFFSet?
<
-25
:
SOURce
:
JITTer
:
RANGe <range>
Command type Sequential command
Parameter
Function
Restriction
<range>
=
CHARACTER PROGRAM DATA>
UI3200
........
3,200 UIPP
UI1600
........
1,600 UIPP
UI800
..........
800 UIPP
U180
............
80 UIPP
U140
............
40
UIPP
UI20
............
20
UIPP
UI05
............
0.5
UIPP
Performs setting of the transmission jitter generation range.
Effective when transmission jitter is generated.
Restrictions on setting apply depending on the Bit Rate.
Example of use
Sets the transmission jitter generation range at 0.5 UIPP.
>
:
SOURce
:
JITTer
:
RAN& UI05
Bit Rate
10
G
UI Range