SOURce
:
TELecom
:
EQUAlizer <boolean>
Command type Sequential command
Parameter
<boolean>
=
<BOOLEAN PROGRAM DATA>
OFF or
0
......
Equalizer OFF
ON
or
1
........
Equalizer
ON
Function
Sets ON/OFF of the Jitter Setting Equalizer.
Example of use
Sets the Jitter Setting Equalizer ON.
>
;
SOURce
:
TELecom
:
EQUAlizer ON
-
--
:
SOURce
:
TELecom
:
EQUAlizer?
Command type Sequential command
Response
<NR
1
NUMERIC RESPONSE DATA>
...................
0
Equalizer OFF
1
...................
Equalizer ON
Function
Makes a query of the setting status of the Jitter Setting Equalizer.
Restriction None
Example of use
>
:
SOURce
:
TELecom
:
EQUAlizer?
<
1
:
SOURce
:
TELecom
:
CLOCk
:
SOURce <csource>
Command type Sequential command
Parameter
<csource>
=
<CHARACTER PROGRAM DATA>
INTernal
.................
Internal
........
LOCK-2MHB 2 MHz (Balanced)
.......
LOCK-2MHU 2 MHz (Unbalanced)
........
LOCK-2MBB 2 Mbit/s (Balanced)
LOCK-2MBU
........
2
Mbit/s (Unbalanced)
LOCK-I SMH
........
1.5 MHz (Unbalanced)
LOCK-1 SMB
........
1.5 Mhit/s (Balanced)
EXTernal
................
External
...........
LOCK-1 OM Lock
1
OM
(Balanced)
Function
Sets the clock source of the transmission signal.
Example of use
Sets the clock source of the transmission signal to internal.
>
:
SOURce
:
TELeconi
:
CLOCk
:
SOURce INTemal