Emerson Process Management CFA-3000 TV Converter Box User Manual


 
10
CFA3000 ISE Analyzer
Instruction Manual
Electrical connections are made to and from the cabinet at the input/output panel, located on the left side of the
analyzer. See Figure 25.
• Terminals are provided for ground, power supply, high/low alarms, output signals, calibration and status
relays
• Output signals  a voltage output of 05 Vdc and a floating, ungrounded current output of 420 mA are
available for remote readout or control of a process.
• Also available are contact closures for the following:
High alarm Calibration
Low Alarm Leak Alarm
To make electrical connections, proceed as follows:
1. Remove solid caps from the holes above the input/output panel by puncturing with an awl and prying
them out. (Note: remove caps only from those holes that will be used.)
2. Replace with permanent connectors.
3. Connect alarms and external computer interfaces (if any).
4. Feed 3wire power cable into analyzer through hole above the input/output panel.
5. Connect wires to appropriate contacts on Terminal Strip TB3.(See next page)
2.6 Prepare Electrical Connections
Use cable glands to hold cables securely in holes above input/output panel.
WARNING:
Connect output signals, if any before connecting the power supply.
CAUTION:
For proper operation and for safety, electrical connections to this equipment must be
made in accordance with local or national electrical code as applicable.
This equipment must be grounded. A qualified electrician should wire this equipment to an
electrical circuit.