Emerson Process Management CFA-3000 TV Converter Box User Manual


 
39
Figure 73. Analog Board DIP Switches
(with #3 set in OFF position for multistream analyzer) Colorimeter Assembly
CFA3000 MANUAL SECTION 7
INSTRUMENT DESCRIPTION
7.3.6. Analog Board DIP Switches
The following list describes the conditions for each switch to be in the ON or OFF position:
Table 79. Analog Board DIP Switch Positions
Switch No. Position Condition Remarks
1 ON
When using Analog Board, Part
# 180B02602
Analog Board has no adjustment
ports for individual output device
and/or 05 Vdc, 420 mA outputs.
1 OFF
When using Analog Board, Part
# 180B02601
Analog Board has individual ports
for adjusting each output device
(up to 6) and/or 05 Vdc,
420 mA outputs.
2 ON Quick calibration  on
2 OFF Quick calibration  off
3 ON
4 ON Nonfunctioning switch
5 ON Nonfunctioning switch
6 ON Nonfunctioning switch
7 ON
For stream ID and “brackets” before
and after calibration
Allows markers to be recorded on
strip charts and other recording
devices.
7 OFF
No markers will be recorded on
strip charts and other recording
devices.
8 ON For 80 column printers Formats printer output
8 OFF For 40 column printers Formats printer output