CFA3000 MANUAL SECTION 3
INSTRUMENT DESCRIPTION
3.3 Set Reagent Supply Time
Use the keypad panel to remind you when you need to replace the reagent. The analyzer will count pump cycles
and count down the number of days left. To specify the duration, proceed as follows:
1. Press once to set counter for 90 days (3 months).
Sample Display: 0 0 9 0
2. Press to exit and clear display.
To determine how many days are left until you have to replace reagent,
Press: once.
Look at the LCD and record the number of days;
then press: again to get out of mode.
3.4 Set Scaling Outputs
Set output 420 mA 05 Vdc from the keypad panel, (refer to Figures 72 and 73), then adjust 420 mA and 05
Vdc at your external device.
Set output scale to: 4 mA/0 V
1. Press: .once then .to display 4 mA and for 0 Vdc.
The sample Display for 4 mA and for 0 Vdc: 0 0 0 4
2. To exit, press
Set output scale to: 20 mA/5 V
1. Press once then to display 20 mA and 5 Vdc.