CFA3000 MANUAL SECTION 3
INSTRUMENT DESCRIPTION
2. Press to simulate one shot of Reagent #4. Look at blue straw to confirm that water is pumped up
through tube. Repeat a few times until you see water is in clear section of tube
3. Press to simulate one shot of Reagent #3. Check green straw and its clear tube
4. Press to simulate one shot of Reagent #2. Check orange straw and its clear tube.
5. Press to simulate one shot of Reagent #1. Check red straw and its clear tube
6. Press to simulate one shot of Sample coming from overflow sampler. Check tube labeled "Sample."
7. Confirm that fluid is going to flowcell.
8. Press to simulate one shot of Standard. Check corresponding clear straw and its clear tube.
9. Press to pump one shot of air into analyzer.
10. Press to simulate one shot of Baseline solution. Check clear straw labeled, BASELINE, and tube.
11. Repeat Step 9 until you see fluid shoot into reaction chamber.
12. Press to simulate evacuation from the reaction chamber to the flowcell. Check the tube going up to
flowcell.
13. Press to simulate one shot of Grab solution. Check clear straw and tube.
14. Press to wash flowcell.
1. Press ZERO to test air. Press ZERO to turn off.