Emerson Process Management CFA-3000 TV Converter Box User Manual


 
20
CFA3000 MANUAL SECTION 3
INSTRUMENT DESCRIPTION
3.2.5 Set Up Stream Parameters
The analyzer is programmed to measure a singlestream system at 4 sample cycles per hour. The multistream
default is 1 cycle per stream, but you may define parameters for individual streams, such as modifying the num
ber of cycles or the alarm threshold values.
The alarm values are used to activate alarms or controllers when the sample is outside of the specified concen
tration range. A high alarm will activate when the sample concentration exceeds a preset limit. A low alarm will
activate when the sample concentration is below the preset limit.
The contacts for the alarm outputs are located at the input/output panel on the cabinet side. The system must be
in an alarm condition for at least one minute to activate the alarm contact.
Use the logical flows below to define desired stream parameters.
Figure 34. Logic Flow for Singlestream Parameter Definition