Emerson Process Management CFA-3000 TV Converter Box User Manual


 
53
CFA3000 MANUAL APPENDIX
APPENDIX B SPECIFICATIONS
Accuracy: All twodecade ranges up to 1000 ppb
maximum: ±10% of reading or ±2 ppb, whichever is
greater
0.110 ppm range = ±10% of reading or ±0.5 ppm,
whichever is greater
Automatic Calibration: 40 minutes
Total Calibration: 40 minutes, twice/week
Reagent Consumption: 5 L of each reagent per 3
months
Standard Consumption: 1 L of high standard,1 L of
low standard per 3 months
Sample Requirements:
Flow: 4050 mL/min
Current standard bottles are:
“0”(baseline): 1.0 L.
Fullscale: 5.0 L
Pressure: <10 psig (<170 kPa) (overflow cup must
drain to atmosphere)
Temperature: 050° C (32122° F)
Filter: 8 micron filter in sample line is
Recommended; (#180134108).
Multiple Stream Option: General: up to 6 sample
streams; measurement is sequential
Response Time: 15 minutes (methoddependent)
Outputs: 420 mA or 05 Vdc (nonisolated).
(Isolated output is optional.)
Load requirements for 420 mA output: 450 ohms
minimum, 1200 ohms? maximum
Load requirements for 05 Vdc output: 10 k Ohms,
minimum
The range of outputs is the same as the measure
ment range. Parallel output for printer is optional for
single stream, standard for multistream monitors.
RS232C output is optional for single stream monitors,
standard for multistream monitors.
Alarms:
General: contact closures for
leak detection,
calibration in progress,
Contact closures are rated:2 A at 250 Vac or 3 A
at 30 Vdc; delay on/off is 1 minute
Streams 2 through 6 (multistream option):
• One alarm relay for each stream; contacts have
dc ratings
• 100 Vdc (maximum  carry only)
• 0.5 A (maximum  switched)
• delay on/off time is 1 minute
Display: Single line LCD, 4 1/2 digits, readout in
concentration units. For multistream option display
remains at last value until new stream reading is stable.
Status LEDs:
• Leak,
• Lamp failure,
• Quarterly maintenance due
• Calibration in progress.
For multistream analyzers.
StatusLEDs show:
• the sample stream currently selected
• the sample stream for which data is being
displayed.
Environmental: The analyzer is not intended for out
door use. Install in a clean, indoor environment.
Ambient Temperature:
40120° F (1045° C) must be regulated to ± 10°
within range, avoid temperature swings
greater than ± 10° F (± 5± C) over a 12hour period
Humidity: 595% (noncondensing)
Pollution degree: Standard
Altitude: Not applicable