CFA3000 MANUAL APPENDIX
Method: Sodium01A
(0.110, 1100, 101000 ppb to 1000 ppm (contact
Rosemount Analytical for other ranges)
General Description:
For this procedure for the determination of sodium,
the sodium ion is measured potentiometrically using
ion selective electrodes . The addition of an ionic
strength adjustor to the sample maintains constant
background ionic strength and pH.
Sodium in Water
Range: 0.110, 1100, 101000 ppb up to 1000 ppm
or higher.
Reagents Kit: (reagent & standards)
185301130
REAGENT #1, Buffering Reagent:
185301134
Working Standards
10 ppb: 185301133
100 ppb: 182301132
Operating Notes
1. The chemicals used for reagents and distilled water
should be free of sodium.
2. The sample stream must be filtered to 10 microns
or less before being introduced to the system with
an inline filter such as Rosemount Analytical' IF
100
(Part #180140001).
APPENDIX A CHEMISTRIES