Teledyne 200A TV Converter Box User Manual


 
Teledyne API Model 200A NO
X
Analyzer Instruction Manual, 02246, Rev. G, DCN 5247
Electric Test (ET) Procedure:
1.
To adjust ET press SETUP-MORE-DIAG, then scroll to ELEC TEST and press ENTR.
2.
Scroll the TEST functions until PMT is displayed.
3.
Adjust R27 until 2000 mV ± 100 is displayed.
4.
Press EXIT to return to SAMPLE mode.
Optic Test (OT) Procedure:
1.
To adjust OT press SETUP-MORE-DIAG, then scroll to OPTIC TEST and press ENTR.
2.
Scroll the TEST functions until PMT is displayed.
3.
Adjust R25 until 2000 mV ± 100 is displayed.
4.
Press EXIT to return to SAMPLE mode.
If this procedure does not produce an instrument that will properly span, please contact your
local distributor or the Teledyne API factory. Teledyne API's phone number is on the front page
of this manual.
9.2 Performance Problems
When the response from a span check is outside the control limits, the cause for the drift should
be determined, and corrective action should be taken. Some of the causes for drift are listed
below:
NOTE
It has been our experience that about 50% of all analyzer performance
problems are sooner or later traced to leaks in some part of the system.
1.
Fluctuations in flow. Such as leaks or plugged orifices.
2.
Lack of preventive maintenance.
3.
Change in zero air source.
A.
Air containing NO leaking into zero air line.
B.
Saturation of charcoal and/or Purafil scrubbers.
9-26
PRINTED DOCUMENTS ARE UNCONTROLLED